Disposable Rf Plasma Electrodes

Products Description
|
Code |
Working Length (mm) |
Plug Type |
|
RPE08-40S |
400±l0mm |
Type S - Rectangle |
|
RPE08-34S |
340±l0mm |
|
|
RPE08-24S |
240±l0mm |
|
|
RPE08-15S |
150±10mm |
|
|
RPE08-40R |
400±l0mm |
Type R-Round |
|
RPE08-34R |
340±l0mm |
|
|
RPE08-24R |
240±l0mm |
|
|
RPE08-15R |
150±l0mm |
|
|
RPE08-40F |
400±l0mm |
Type F-Split |
|
RPE08-34F |
340±l0mm |
|
|
RPE08-24F |
240±l0mm |
|
|
RPE08-15F |
150±l0mm |
Contraindication
1) Disposable radiofrequency plasma electrode is forbidden to be used in any operation without any conductive fluid.
2) Disposable radiofrequency plasma electrode is forbidden to be used for the patients with cardiac pacemaker or other electronic devices.
3)There may be therapeutic damage to surrounding tissue due to electrosurgery.
Product Sterilization Information
Disposable Rf Plasma Electrodes is sterilized by Radiation. Do not re-sterilize
Validity
3 years.
Mateched HF
Cutting: 1600Vp (3200vpp); Coagulation: 2000vp (4000vp-p).
Features
Complete ablation and hemostasis in all directions under transforaminal endoscope, without damaging tissue and nerves
Controllable power for precise surgery
Precise ablation of nucleus pulposus tissue within the intervertebral disc
Short operation time and less pain for the patient
No sequelae, low recurrence rate
Structure and Component
The product consists of bipolar RF plasma electrode blade with guide wire, sleeve, double lumen catheter, insulated handle seat, insulated handle slider and cable. The product is disposable, sterilized by electron accelerator irradiation, and sterile, valid for three years.

1. Bipolar electrode blade with guide wire 2. Sleeve 3. Insulated handle seat 4. Insulated handle slider 5. Cable with plug
Figure 1 Product Structure Diagram
1. Mode and power output table:
|
Mode |
Frequency (MHz) |
Power output (W) at impedance 500Ω |
|
Cutting |
4.0 |
100 |
|
Cutting/Coagulation |
4.0 |
70 |
|
Hemostasis |
4.0 |
50 |
|
Ablation |
4.0 |
50 |
|
Bipolar hemostasis |
1.7 |
40 |
2. Output voltage: AC 0-360V
3. Function:
The RF plasma surgical treatment system suitable for electrodes should have the following functions:
(A) Function of setting ablation/cutting, coagulation/hemostasis modes;
(B) Function of connecting and displaying electrodes (blade and needles);
(C) Function of indicating the working state;
(D) Function of output power regulation.
4. Interface:
Bipolar socket hole diameter: Φ3 ± 2mm, hole spacing: 16 ± 2mm, as shown below.